3500 Series Continuous 2 Mode

$182.51$259.36

SKU: 3500 Series Continuous 2 Mode

Image may not reflect product configuration.

The unique modular design makes repair a breeze for any biomed or central supply department. The unit is made up of six (6) easy-to-replace modular components.

Unlike the normal plastic inlet and outlet connections on most units, the cast metal backplate connections are extremely durable, thereby saving unnecessary repair costs.

As a result of its modular design, the unit is 25-50% smaller than other suction regulators making it the product of choice for crowded wall and rail areas.

The case is made of strong plastic. It features a smooth surface with rounded corners making cleaning easy. The case also houses the gauge for extra protection.

The large control knob and easy-to-read settings make vacuum adjustments simple, even in emergency situations.

The Pediatric vacuum regulator is restricted to 170 mmHg (± 10 mmHg) to prevent high vacuum situations and the gauge face is imprinted with the Precision Medical’s “Stork Symbol” to identify the unit for pediatric applications.

Regulator Mode: Continuous 2 Mode (0-150 mmHg)

Contact Precision Medical for MR Conditional options. Custom colors are available. Please call customer service at 1-800-272-7285 for details.

   Caution! U.S. Federal law restricts this device to sale by or on the order of a physician.

Dimensions
Length: 3.7 in (9.4 cm)
Width: 2.8 in (7.1 cm)
Height: 5.2 in (13.2 cm)
Weight: 1.1 lbs (.50 kg)
Gauge Range
PM3500: 0 – 150 mmHg
*PM3500E: 0 – 200 mmHg (0 – 26 kPa)
*Counterclockwise direction
Gauge Accuracy
Analog: ± 5% of MAX

Digital Gauge: ± 1% of Full Scale
Analog Gauge: ± 5% of MAX within ref. Indicator
Vacuum Ports
1/8″ NPT Female
Modes of Operation
OFF: No Vacuum

REG. (regulated): provides an adjustable, continuous vacuum level
Flow
REG:51 l/min
LINE: 55 l/min
Note: maximum flow is obtained with a vacuum source of 21″ Hg (71.1 kPa)
Maximum Vacuum
PM3500D: Restricted to 170 mmHg (23 kPa) ± 10 mmHg (1.3 kPa)

PM3500E: Restricted to 160 mmHg (21 kPa) ± 10 mmHg (1.3 kPa)
Operating Environmental Limits
0°F to 122°F (-18°C to 50°C)
Recommended Environmental Operating Limits
55°F to 85°F (13°C to 29°C)
Storage Environmental Limits
Temperature Range: -4°F to 140°F (-20°C to 60°C)

Humidity: Max 95% Non-condensing
Battery
Digital Vacuum Gauge Models ONLY
3 Volt Lithium, ½ AA
*Specifications are subject to change without prior notice.