Air-Oxygen Blender

It is the only unit available that allows the clinician to turn the bleed control to ON or OFF providing quieter operation. This feature is especially useful in the NICU and PICU.

The tough ABS housing makes the unit durable and the selection control knob is recessed to prevent accidental damage.

The unit is available in two models, high and low flow. The blender is easily mounted with a universal mounting bracket. It can be detached from the mounting bracket without removing hoses.

The unique modular design can cut routine maintenance time in half as well as reduce maintenance costs. The unit has easy-to-replace modular components and is easily maintained by all biomed departments.

A large selection control knob makes oxygen percentage adjustments easy. The infection-control friendly housing is smooth and easy to clean.

The highly accurate unit maintains FI02 even at low flows. It is economical and provides cost-savings with less waste.

For custom outlet configurations and assembly sizes, please contact customer service by using this form or by calling 1-800-272-7285.

Retail price

$792.00

$968.00

$878.24

$1,054.24

$99,999,999,999.00 - $0.00

SKU

*Items with custom pricing will be processed by our customer service representative once your order has been submitted.

Configuration selection does not change product image.

Available Accessories

Specifications

Primary Outlet Flow Range

PM5200 High FlowPM5300 Low Flow
15 - 120 L/min 3 - 30 L/min
supply pressures at 50 psi (3.4 bar) with BLEED closed supply pressures at 50 psi (3.4 bar) with BLEED closed

Auxiliary Outlet Flow Range

PM5200 High FlowPM5300 Low Flow
2 - 100 L/min 0 - 30 L/min
supply pressures at 50 psi (3.4 bar) with BLEED closed supply pressures at 50 psi (3.4 bar) with BLEED closed

Bleed Flow

PM5200 High FlowPM5300 Low Flow
13 L/min or less at 50 psi (3.4 bar) 3 L/min or less at 50 psi (3.4)

Maximum Combined Flow (All Outlets)

PM5200 High FlowPM5300 Low Flow
≥ 120 L/min ≥ 30 L/min

Bypass Flow (Loss of Air or Oxygen Supply)

PM5200 High FlowPM5300 Low Flow
≥ 85 L/min ≥ 45 L/min

Bypass Alarm Activation

PM5200 High Flow 50 psi(3.45 bar)PM5200 High Flow 60 psi(4.14 bar)PM5300 Low Flow 50 psi(3.45 bar)PM5300 Low Flow 60 psi(4.14 bar)
13 - 25 psi 16 - 24 psi 18 - 22 psi 16 -24 psi
0.9 - 1.7 bar 1.1 - 1.65 bar 1.2 - 1.5 bar 1.1 - 1.65 bar

Alarm Reset

When pressure differential is 6 psi (0.4 bar) or less

Alarm Sound Level

≥ to 80 db at 1 ft (0.3 m)

Oxygen Concentration Adjustment Range

21 - 100%

Gas Supply Pressure

30 - 75 psi (2.1 - 5.2 bar) Air and Oxygen within 10 psi (0.69 bar) of each other

Mixed Gas Stability

± 1% Oxygen

Connection Types

DISS Type - Air & Oxygen Inlets & Outlets and / or NIST Type - Air & Oxygen Inlets

Dimensions:(without fittings)

Depth: 4.9 in (12.5 cm)
Width: 2.3 in (5.7 cm)
Height: 4.1 in (10.4 cm)

Weight

Unit Weight: 2.29 lbs (1.04 kg)
Shipping Weight: 2.95 lbs (1.34 kg)

Operating Temperature Range

59°F to 104°F (15°C to 40°C)

Transport / Storage Requirements

Temperature Range: -10°F to 140°F (-23°C to 60°C)
Humidity: Max 95% Noncondensing

FI02 Accuracy

± 3% of full scale

Pressure Drop

Low Flow: ≤ 2 psi (0.14 bar) at inlet pressure from 30 - 90 psi (2.1 - 6.2 bar) and at 10 L/min flow rate at 60% FI0²
High Flow: ≤ 3 psi (0.21 bar) at inlet pressure from 30 - 90 psi (2.1 - 6.2 bar) and at 30 L/min flow rate at 60% FI0²

Pressure Drop Additional Info

The Air-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.
The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 21647.

Dryness and Composition for inlet gases

Air: Medical Air supply should meet the requirements of ANSI Z86.1 - 1973 commodity specification for Air, type 1 grade D or better.
Oxygen: Oxygen supply must meet all requirements of USP Medical Grade Oxygen.
Dew Point(ONLY for CE requirements: Both inlets should remain 10°F (5.55°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F (-3.9°C) and a pressure of 90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.
Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 L/min for the high flow Blender, and 3 L/min for the low flow Blender.

*Specifications are subject to change without prior notice.*