Dial Flowmeter
The lack of a flow tube and the durable metal body prevent the breakage problems long associated with standard flowmeters.
The compact size allows the unit to fit into tight headwall areas and makes it virtually damage-proof when other equipment is installed.
Accuracy is assured at every flow setting without the need for a visible float as in thorpe tube flowmeters.
The knurled flow adjustment knob and the large easy-to-read numbers make flow adjustment easy.
The unit is available for headwall outlets as well as for ceiling mount drops. It comes in two different flow ranges: .5-15 L/min with flush and 1-25 L/min (both available in oxygen and air).
For price inquiries, please use this form.
Specifications
Flow Range
Flow Range | Settings (lpm) | Pressure |
---|---|---|
0 - 15 L/min | Off, .5, 1, 2, 3, 4, 5, 6, 8, 10, 15, and Flush | 50 psi |
0 - 25 L/min | Off, 1, 2, 3, 4, 6, 8, 10, 12, 15, 20, and 25 | 50 psi |
NOTE: There is no flow between settings. |
Max Flush Flow
Model | Gas | Max Flush Flow | Increments (L/min) |
---|---|---|---|
7MFA1001 7MFA2001 |
Oxygen Air |
40 - 60 L/min | .5, 1, 2, 3, 4, 5, 6, 8, 10, 15 Flush |
7MFA1001TC 7MFA2001TC |
Oxygen Air |
40 - 60 L/min | .5, 1, 2, 3, 4, 5, 6, 8, 10, 15 Flush |
7MFA1101 7MFA2101 |
Oxygen Air |
N/A | 1, 2, 3, 4, 6, 8, 10, 12, 15, 20, 25 |
7MFA1101TC 7MFA2101TC |
Oxygen Air |
N/A | 1, 2, 3, 4, 6, 8, 10, 12, 15, 20, 25 |
Flowmeters calibrated at 50psi (3.4 bar), 70°F (21°C0). standard atomspheric pressure. |
Accuracy
± 10% of the indicated flow |
Operating Temperature
69°F to 71°F (20.6°C to 21.6°C) |
Storage Temperature
-40°F to 140°F (-40°C to 60°C) |
Max Humidity
95% |
Storage/Transport outside the specified range may cause damage to the flowmeter. |