Helium Oxygen Blender
The HeliO2 Helium-Oxygen Blender is the only freestanding pneumatic HeliO2 blending device on the market today. The ability to turn the bleed control collar ON or OFF provides ± 3% FiO2 accuracy as well as quieter operation and maximal heliox tank durations.
The tough housing makes the unit durable and the selection control knob is recessed to prevent inadvertent changes in FiO2. The modular design can reduce maintenance time and costs. The unit can be easily maintained by all biomed departments.
The unit is available in four models, High-Flow 80/20 or 70/30 and Low-Flow 80/20 or 70/30. When paired with the optional cylinder clamp and pole, clinician setup time is greatly reduced and tank changes are easy. Multiple accessories can conform to clinician and user needs.
Features
The world's only freestanding pneumatic helium oxygen blender
Modular and customizable
Multiple accessories available for suiting various clinical application
For price inquiries, please use this form.
Specifications
Primary Outlet Flow Range
High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox |
---|---|
15 - 120 L/min | 3 - 30 L/min |
supply pressures at 50 psi (3.4 bar) with BLEED Closed | supply pressures at 50 psi (3.4 bar) with BLEED Closed |
Auxiliary Outlet Flow Range
High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox |
---|---|
2 - 100 L/min | 0 - 30 L/min |
supply pressures at 50 psi (3.4 bar) with BLEED Open | supply pressures at 50 psi (3.4 bar) with BLEED Open |
Bleed Flow @ 100% FI02
High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox |
---|---|
13 L/min or less at 50 psi (3.4 bar) | 3 L/min or less at 50 psi (3.4 bar) |
Maximum Combined Flow (All Outlets)
High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox |
---|---|
≥ 120 L/min | ≥ 30 L/min |
Bypass Flow (Loss of Helium or Oxygen supply)
High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox |
---|---|
≥ 85 L/min | ≥ 45 L/min |
Bypass Alarm Activation
High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox - 50 psi (3.45 bar) | High Flow PM5580: 80/20 Heliox PM5570: 70/30 Heliox - 60 psi (4.14 bar) | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox - 50 psi (3.45 bar) | Low Flow PM5480: 80/20 Heliox PM5470: 70/30 Heliox - 60 psi (4.14 bar) |
---|---|---|---|
13 - 25 psi | 16 - 24 psi | 18 - 22 psi | 16 - 24 psi |
0.9 - 1.7 bar | 1.1 - 1.5 bar | 1.2 - 1.5 bar | 1.1 - 1.65 bar |
Alarm Reset
When pressure differential is 6 psi (0.4 bar) or less. |
Alarm Sound Level
≥ to 80 db at 1 ft (0.3 m) |
Oxygen Concentration Adjustment Range
PM5480 / PM5580 | PM5470 / PM5570 |
---|---|
20 - 100% | 30 - 100% |
Gas Supply Pressure
30 - 75 psi (2.1 - 5.2 bar) Heliox & Oxygen within 10 psi (0.69 bar) of each other |
Mixed Gas Stability
± 1% Oxygen |
Connection Types
DISS Type - Heliox & Oxygen Inlets & Outlets and / or NIST Type - Heliox & Oxygen Inlets Note: All flow - rate values are as measured from an Oxygen flowmeter (uncorrected). |
Dimensions (without fittings)
Depth: 4.9 in (12.5 cm) |
Width: 2.3 in (5.7 cm) |
Height: 4.1 in (10.4 cm) |
Weight
Unit Weight: 2.29 lbs (1.04 kg) |
Shipping Weight: 2.95 lbs (1.34 kg) |
Operating Temperature Range
59°F to 104°F (15°C to 40°C) |
Transport / Storage Requirements
Temperature Range: -10°F to 140°F (-23°C to 60°C) |
Humidity: Max 95% Concondensing |
FI0² Accuracy:* ± 3% of full scale @ 50 psi (3.4 bar) |
Pressure Drop
Low Flow: ≤ 2 psi (0.14 bar)at inlet pressures from 30 - 90 psi (2.1 - 6.2 bar) and at 10 L/min flow rate at 60% FI02. |
High Flow: ≤ 3 psi (0.14 bar)at inlet pressures from 30 - 90 psi (2.1 - 6.2 bar) and at 10 L/min flow rate at 60% FI02. |
Pressure Drop Additional Info
The Helium - Oxygen Blender has been cleaned for Oxygen Service prior to delivery. |
The Helium - Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195. |
The Oxygen Analyzer should comply with ISO 21647. |
Dryness and Composition for inlet gases
Heliox: Medical grade 80/20 or 70/30 required. |
Oxygen: Oxygen supply must meet all requirements of USP Medical Grade Oxygen. |
Dew Point (ONLY for CE requirements: Both inlets should remain 10°F (5.55°C) or more below the lowest temperature to which the air distribution system equipment is exposed. At a temperature of 25°F (-3.9°C) and a pressure of 90 psi (6.33 kg/cm2) this equates to 2000 mg/m3. |
Accuracy of FI0² will be affected if bleed flow is not engaged at low flows. (At or below 3 L/min for Low Flow and 15 L/min for High Flow). |
When Heliox tank pressure and oxygen outlet pressures are unbalanced, bleed may need to be engaged at a higher liter flow to maintain accuracy. |
*Specifications are subject to change without prior notice.*